MW-RSP-0500-24

MeanWell RPS-500-24 500W 24V Medical-Grade PCB Power Supply
Overview: The MeanWell RPS-500-24 is a 24 VDC, 500 W-class medical-grade AC/DC power supply in a compact 5"×3" PCB form factor with 2×MOPP primary-to-secondary isolation. It is built for medical device manufacturers, industrial automation panels, and ITE/telecom equipment requiring a regulated 24 VDC rail.
Description:
The RPS-500 series is an active-PFC AC/DC converter that operates from an 80264 VAC input (113370 VDC) and delivers a power factor above 0.94 at full load. The 24 V model provides 13.4 A / 321.6 W continuously under natural convection or 20.8 A / 499.2 W with 25 CFM forced airflow, and can supply 550 W of peak power for 3 seconds. The main output is adjustable from 22.8 to 25.2 VDC with 200 mVp-p maximum ripple, offering output quality suitable for powering sensitive analog and digital stages. The PCB-type chassis measures 127 × 76.2 × 41 mm, minimizing footprint in embedded device designs.
On the isolation side, the unit provides 2×MOPP primary-to-secondary separation (I/P-O/P 4 kVAC, I/P-FG 2 kVAC, O/P-FG 1.5 kVAC) and complies with IEC/EN 60601-1, ANSI/AAMI ES60601-1, CAN/CSA C22.2 No. 60601-1, and IEC/EN/UL 62368-1. For EMC, it meets BS EN/EN55011 and BS EN/EN55032 emission standards together with BS EN/EN55035 and BS EN/EN60601-1-2 immunity standards, with separate EMI performance defined for Class I (with FG) and Class II (without FG) configurations. Immunity is verified at 4 kV line-to-FG / 2 kV line-to-line surge, 15 kV air / 8 kV contact ESD, and 10 V/m RF field levels. Per the manufacturer's note, EMC tests must be re-confirmed in the final equipment with the unit mounted on a 360×360 mm, 1 mm thick metal plate.
For control and monitoring, the unit offers remote on/off via PS-ON, Power Good / Power Fail TTL signals, and a remote sense line. A 5 VDC standby output (0.6 A without fan, 1 A with a 25 CFM fan) and a 12 VDC / 0.5 A auxiliary output for fan supply reduce the need for external DC/DC stages in system designs. Short circuit, overload (105~135%), over voltage, and over temperature protections, combined with a -30 ~ +70 °C operating range and 4000 m operating altitude, support reliable operation in field-type installations. MTBF is rated at 1132.3K hours (Telcordia SR-332), and the unit carries a 3-year warranty.
Key Highlights:
- 24 VDC main output; 20.8 A / 499.2 W continuous with 25 CFM airflow, 13.4 A / 321.6 W under natural convection
- 550 W peak power capability for 3 seconds
- Wide 80
264 VAC and 113370 VDC input range with >0.94 power factor at full load - 2×MOPP primary-to-secondary isolation, 4 kVAC withstand voltage; IEC/EN 60601-1 and IEC/EN 62368-1 compliance
- PS-ON, Power Good / Power Fail, and remote sense; 5 VDC standby plus 12 VDC fan supply
- 127 × 76.2 × 41 mm PCB form factor, -30 ~ +70 °C operating range, 3-year warranty
Technical Specifications:
| Parameter | Value |
|---|---|
| Manufacturer | MeanWell |
| Model | RPS-500-24 |
| Product Group | Power Supplies |
| Output Voltage | 24 VDC |
| Rated Current (25 CFM airflow) | 20.8 A |
| Rated Current (natural convection) | 13.4 A |
| Rated Power | 499.2 W (forced air) / 321.6 W (convection) |
| Peak Power | 550 W (3 sec) |
| Voltage Adjustment Range | 22.8 ~ 25.2 VDC |
| Ripple & Noise (max.) | 200 mVp-p |
| Input Voltage | 80 ~ 264 VAC / 113 ~ 370 VDC |
| Input Frequency | 47 ~ 63 Hz |
| AC Input Current (typ.) | 5.8 A @115 VAC, 2.9 A @230 VAC |
| Power Factor | >0.94 @230 VAC, >0.98 @115 VAC (full load) |
| Inrush Current (typ.) | 40 A @115 VAC, 80 A @230 VAC (cold start) |
| Efficiency (typ.) | Up to 94% |
| Standby Output | 5 VDC @ 0.6 A (no fan) / 1 A (25 CFM fan) |
| Fan Supply | 12 VDC @ 0.5 A |
| Control & Monitoring | PS-ON, Power Good / Power Fail, remote sense |
| Protections | Short circuit, overload, over voltage, over temperature |
| Operating Temperature | -30 ~ +70 °C |
| Storage Temperature | -40 ~ +85 °C |
| Operating Humidity | 20 ~ 90% RH non-condensing |
| Operating Altitude | Up to 4000 m |
| Withstand Voltage | I/P-O/P 4 kVAC, I/P-FG 2 kVAC, O/P-FG 1.5 kVAC |
| Isolation | Primary-Secondary 2×MOPP |
| Safety Standards | IEC/EN 60601-1, ANSI/AAMI ES60601-1, CAN/CSA C22.2 No. 60601-1, IEC/EN/UL 62368-1, EAC TP TC 004 |
| EMC | BS EN/EN55011, BS EN/EN55032, BS EN/EN55035, BS EN/EN60601-1-2 |
| MTBF | 1132.3K hrs (Telcordia SR-332), 144.2K hrs (MIL-HDBK-217F, 25 °C) |
| Dimensions | 127 × 76.2 × 41 mm |
| Weight | 0.46 kg |
| Warranty | 3 years |